摘要: |
目的 分析细菌内毒素凝胶法检查中出现的不同结果。方法 采用细菌内毒素动态浊度法考察样品对鲎试剂的干扰作用。结果 不同倍数的样品稀释液存在抑制和增强作用。结论 细菌内毒素动态浊度法能更准确地反映样品的干扰情况 |
关键词: 细菌内毒素 动态浊度法 凝胶法 干扰试验 |
DOI: |
分类号: |
基金项目: |
|
Analysis of the various results in gel-clot technique by kinetic turbidimetric technique |
ZHANG Jin-song
|
Abstract: |
OBJECTIVE To analyze the various results in gel-clot technique assay.METHOD Kinetic turbidimetric technique assay was used to evaluate the interference of sample to TAL.RESULTS The result showed the inhibition and enhancement of sample to TAL in variou |
Key words: bacterial endotoxin kinetic turbidimetric technique gel-clot technique interferential test |