• 首页期刊简介编委会刊物订阅专栏专刊电子刊学术动态联系我们English
引用本文:何家伟,汪洋,纪保超,高惠静,杨建华.HPLC测定人工关节假体感染患者关节液中万古霉素浓度[J].中国现代应用药学,2019,36(23):2949-2952.
HE Jiawei,WANG Yang,JI Baochao,GAO Huijing,YANG Jianhua.Determination of Vancomycin Concentration in Synovial Fluid of Patients with Prosthetic Joint Infection by HPLC[J].Chin J Mod Appl Pharm(中国现代应用药学),2019,36(23):2949-2952.
【打印本页】   【HTML】   【下载PDF全文】   查看/发表评论  【EndNote】   【RefMan】   【BibTex】
←前一篇|后一篇→ 过刊浏览    高级检索
本文已被:浏览 1986次   下载 913 本文二维码信息
码上扫一扫!
分享到: 微信 更多
HPLC测定人工关节假体感染患者关节液中万古霉素浓度
何家伟1, 汪洋2, 纪保超2, 高惠静1, 杨建华1
1.新疆医科大学第一附属医院药学部, 乌鲁木齐 830054;2.新疆医科大学第一附属医院关节外科, 乌鲁木齐 830054
摘要:
目的 建立HPLC测定人工关节假体周围感染患者关节液中万古霉素浓度的方法。方法 以替硝唑为内标,采用Hypersil C18柱(250 mm×4.6 mm,5 μm),流动相为乙腈-0.012 5mol·L-1磷酸二氢钾缓冲液(pH=3.2)(10:90),流速1 mL·min-1,柱温35℃,检测波长236 nm,进样量10 μL。结果 关节液中万古霉素浓度在0.906 8~90.681 8 μg·mL-1线性良好。批内、批间精密度RSD均<15%,准确度85.89%~103.51%。结论 本法易于操作,精密度和准确度良好,适用于人工关节假体周围感染患者关节液中的万古霉素浓度监测。
关键词:  万古霉素  高效液相色谱法  人工关节假体周围感染
DOI:10.13748/j.cnki.issn1007-7693.2019.23.013
分类号:R927.2
基金项目:
Determination of Vancomycin Concentration in Synovial Fluid of Patients with Prosthetic Joint Infection by HPLC
HE Jiawei1, WANG Yang2, JI Baochao2, GAO Huijing1, YANG Jianhua1
1.Department of Pharmacy, The First Teaching and Affiliated Hospital of Xinjiang Medical University, Urumqi 830054, China;2.Department of Joint Surgery, The First Teaching and Affiliated Hospital of Xinjiang Medical University, Urumqi 830054, China
Abstract:
OBJECTIVE To establish a method for determination of vancomycin concentration in synovial fluid of patients with prosthetic joint infection by HPLC. METHODS Used tinidazole as internal standard substance, vancomycin concentration in synovial fluid was analyzed by Hypersil C18(250 mm×4.6 mm, 5 μm) with mobile phase of acetonitrile-0.012 5 mol·L-1 potassium dihydrogen phosphate buffer solution(pH=3.2)(10:90). The flow rate was 1 mL·min-1. The column temperature was 35℃ and the injection volume was 10 μL. RESULTS Vancomycin concentration in synovial fluid was in well linearity in the range of 0.906 8-90.681 8 μg·mL-1. The relative standard deviations of intra-batch and inter-batch precision was <15%, and the accuracy was between 85.89%-103.51%. CONCLUSION The method is simple, sensitive and accurate, which is suitable for monitoring of vancomycin concentration in synovial fluid of patients with prosthetic joint infection.
Key words:  vancomycin  HPLC  prosthetic joint infection
扫一扫关注本刊微信