引用本文: | 王文强,陈彬彬,吴海龙,章家新,谢加奖.LC-MS/MS测定人血浆中哌甲酯[J].中国现代应用药学,2010,27(10):942-944. |
| .LC-MS/MS Determination of Methylphenidate in Human Plasma[J].Chin J Mod Appl Pharm(中国现代应用药学),2010,27(10):942-944. |
|
|
摘要: |
目的 建立LC-MS/MS测定人血浆中盐酸哌甲酯的浓度。方法 待测血浆1.0 mL,经1 mol·L-1的氢氧化钠40 μL碱化后,用4 mL乙醚萃取处理,采用Eclipse XDB-C18 (4.6 mm×150 mm,5 μm)色谱柱,以甲醇和水为流动相梯度洗脱,流速1.0 mL·min-1。以电喷雾正离子源离子化,检测离子对盐酸哌甲酯为234.0/84.0,内标卡马西平为237.0/194.0。结果 血浆中内源性物质对测定无干扰,最低定量限为0.1 ng·mL-1,在1~100 ng·mL-1内盐酸哌甲酯线性关系良好(r=0.999 1),日内、日间RSD均小于8%,样品分析时间为10 min。结论 该法专属性强、灵敏度高,操作简便快速,测定结果可靠,适于进行盐酸哌甲酯血药浓度监测。 |
关键词: LC-MS/MS 盐酸哌甲酯 药动学 |
DOI: |
分类号: |
基金项目: |
|
LC-MS/MS Determination of Methylphenidate in Human Plasma |
WANG Wenqiang CHEN Binbin WU Hailong ZHANG Jiaxin XIE Jiajiang
|
Abstract: |
OBJECTIVE To establish an LC-MS/MS method for determination of methylphenidate in human plasma. METHODS 1.0 mL plasma was basified with 40 μL 2 mol·L-1 NaOH and extracted with 4 mL ether. The separation was performed on an Eclipse XDB-C18 column (150 mm×4.6 mm, 5 μm). The mobile phase consisted of methanol and water for gradient elution. Methylphenidate was determined by LC-MS/MS operated in the positive ion mode. RESULTS The limit of quantitation was 0.1 ng·mL-1, and a good linearity was obtained in the range of 1–100 ng·mL-1(r=0.999 1). The RSD of inter-day and intra-day precision were less than 8%. Each analysis was completed within 10 min. CONCLUSION The method is accuracy, sensitive and simple, proved to be suitable for the study of therapeutic drug monitoring of methylphenidate. |
Key words: LC-MS/MS methylphenidate hydrochloride pharmacokinetics |